Mechanical Ventilator For corona Virus Pademic.
Efficient
The ELICIAE MV20 utilizes a Venturi valve system, which supplies sufficient and optimal tidal volumes on patients with reduced and/or recovering pulmonary function.
Optimal handling
On the ELICIAE MV20 ventilator, function keys and displayed parameters are optimized for easy setup of respiratory parameters and quick start the ventilator, yet still ensures safety and respiratory management for the patient.
Safety
The ELICIAE MV20 ventilator offers a variety of breathing modes, from complete mechanical ventilation in patients with severe pulmonary dysfunction, to mechanical ventilation in patients with gradually recovered pulmonary function. Back up mode is the suitable therapy to rehabilitate patients who have gradually recovered pulmonary function and have spontaneous breathing cycle.
Efficiently reduce cross-infection
The ELICIAE MV20 system combines heat and moisture exchangers with bacterial filters (HME), together with a closed suction system to significantly reduce the chance of cross infection for patients and medical personnel. The patient’s expiratory air flow is able to be treated prior to discharge into an environment outside of the patient room to minimize potential cross infection to other patients and medical personnel.
Ventilator Specifications
Mode | Mandatory Ventilation: time cycled pressure-controlled ventilation. Back-up ventilation: flexible operation in back-up mode where CPAP pressure is applied during spontaneous breathing, and mandatory ventilation is acrtivated when APNEA is detected. Monitor: monitor the pulmonary function of the patient as well as to supply oxygen and PEEP. Standby |
Settings & Range
MV baseflow | 3 to 20 Lpm |
APNEA | 1 to 60 sec |
Oxygen Saturation | 21 to 100% |
Inspiration Pressure | 0 to 80 cmH2O |
Inspiration Time | 0.1 to 3.0 sec |
PEEP/CPAP | 0 to 30 cmH2O |
Breath Rate | 1 to 250 bpm |
Alarms

Monitoring
Monitoring | Tidal Volume: 30mL ~ 2,000mL ±10%+10mL Peak Inspiratory Pressure: 0 ~ 80cmH2O ±10% PEE/CPAP: 0 ~ 20cmH2O ±10% Airway pressure: -10 ~ 80cmH2O ±2cmH2O Apnea timeout: 1 ~ 60 second |
Operating Conditions
Operating pressure | 0.3~0.5Mpa |
Electric shock protection class | Class I |
Elictric shock protecion | B type |
IP protection class | IPX0 |
Dimensions | 260mm(W)×190mm(H)×283mm(D) |
Weight | 6kg |
Input power | AC100 to 240V (50 to 60Hz) |
Ambient temperature range | 5 to 40 °C |
Ambient humidity range | 0 to 95% (without condensation) |